MEASURING RANGE
Parameter Sensor Measuring range Resolution Accuracy* T
90
response time
Long-life O2 Electro-chemical From 0 % to 21 % 0.1 % vol. ±0.2 % vol. 30 s
CO
(with H2 compensation) Electro-chemical From 0 to 8000 ppm 1 ppm
From 0 to 200 ppm: ±10 ppm
From 201 to 2000 ppm: ±5% of measured value
From 2001 to 8000 ppm: ±10% of measured value
30 s
Flue gas temperature K thermocouple From -100 to +1250 °C 0.1 °C ±0.4% of measured value or ±1.1 °C 45 s
Ambient temperature Internal NTC From -20 to +120 °C 0.1 °C ±0.5 °C –
Ambient temperature Pt100 (external probe
1/3 DIN) From -50 to +250 °C 0.1 °C ±0.3% of measured value or ±0.25 °C 30 s
Dew-point temperature Calculated** From 0 to +99 °Ctd 0.1 °C – –
DHW temperature TcK (external probe) From -200 to +1300 °C 0.1 °C ±0.4% of measured value or ±1.1 °C –
Differential pressure
Semiconductor
From -20 000 to
+20 000 Pa 1 Pa From -20 000 to -751 Pa: ±0.5 % of measured value ±4.5 Pa
From -750 to -61 Pa: ±0.9 % of measured value ±1.5 Pa
From -60 to 60 Pa: ±2 Pa
From 61 to 750 Pa: ±0.9 % of measured value ±1.5 Pa
From 751 to 20 000 Pa: ±0.5 % of measured value + 4.5 Pa
–
Draft From -10 to +10 Pa
From -1000 to +1000 Pa
0.1 Pa
1 Pa
Losses Calculated** From 0 to 100 % 0.1 % – –
Excess air (λ) Calculated** From 1 to 9.99 0.01 – –
Lower efficiency (ηs) Calculated** From 0 to 100 % 0.1 % – –
Higher efficiency (ηt)
(condensing) Calculated** From 0 to 120 % 0.1 % – –
Opacity index External instrument From 0 to 9 – – –
*All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with required compensation.
Combustion gas analyser KIGAZ 110
2 Go memory (100 000 measurements)
● Step by step procedure menu (gas flow, inspections)
● Self-test menu
● Built-in printer